Display In The Operating Display With An Active Metering Controller; Configuration / Metering Monitoring - Ecolab Elados EMP IV E60 Notice Technique

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

10.3.12.2 Display in the operating display with an active metering controller

Fig. 10.30
Fig. 10.31
Fig. 10.32
Fig. 10.33
NOTE

10.3.13 Configuration / Metering monitoring

If the metering monitoring function is enabled, the pump strokes are compared to the
incoming pulses using a connected external metering monitor (e.g. oval gear meter). If a
configurable tolerance threshold is breached, an alarm is triggered.
If an oval gear meter is connected as a metering monitoring unit and should be
correspondingly evaluated, then the configuration/oval gear meter function must be
NOTE
enabled in addition to the configuration/metering monitoring function
(see chapter 10.3.11).
10.3.13.1 Selecting
417101799_EMP_IV_E60.docx
▼ = Downward metering divergence
The pump is currently delivering less than the preset stroke
frequency (here 100/min for example) requires.
Metering frequency is increased automatically until the actual
value matches the target value.
- = No metering divergence
The pump delivers exactly the preset flow (here 100/min for
example).
No correction is necessary.
▲ = Upward metering divergence
The pump is currently delivering more than the preset stroke
frequency (here 100/min for example) requires.
Metering frequency is reduced automatically until the actual
value matches the target value.
Dongle plate mounted.
Oval gear meter OGM
- 99 -
PLUS
mounted.
Fig. 10.34
-
metering monitoring is inactive.
(Default setting)
 metering monitoring is active.
On activating metering monitoring, a
submenu appears with the following
menu items:
Menu item
see Chapter
• Stop pump
• Strokes
• Deviation
10.3.13.2
10.3.13.3
10.3.13.4
Rev. 9-06.11

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Elados emp iv e60plus

Table des Matières