4.5.3 CALIBRATION OF THE FLOW ZERO POINT

Fill the pipe with the measured fluid and stop the flow. To calibrate the flow zero point,
press 'enter' when CALIB 0 is diplayed within the test menu and select CALIB Y. After
selection the transmitter will automatically set the flow zero-point after 12 seconds.
The sensor must be immersed in fluid 24hrs before calibration. Ensure
there are no air bubbles in the pipe and the fluid is not moving before
commencing the calibration.
This calibration is only valid for the actual parameters (pipe, fitting and fluid
characteristics) and must be completed before the determination of the
K-Factor via the Teach-In method.
CALIB 0
FLOW
FLOW

4.5.4 FLOW SIMULATION

A flow value can be simulated within this menu, allowing the user to test the system without
any liquid being present. The simulated value influences all the outputs including the
relays and pulse output.