3.2.

Turning ON and Fresh Air Setup

Instrument operation is dialog driven from the display with the aid of the
three Function buttons (see Section 2.2.1).
For more information, see the flow charts in Section 7.
Turn the instrument ON with the
The instrument performs a self test:
During the self test, the instrument checks alarm LEDs, audible alarm
vibrating alarm and for missing sensors.
In case of a missing sensor, the instrument displays the Sensor
Missing screen (see 2.5) and alarms until it is turned OFF.
Otherwise, the turn-ON sequence continues.
The instrument displays:
Startup logo
Software version, instrument name and serial number, company name,
department and user names
Sampling safety test (if equipped with an integral pump)
Combustible gas type and sensor units (monochrome display only)
Alarm setpoints Low Alarm
Alarm setpoints High Alarm
Alarm setpoints STEL Alarm
Alarm setpoints TWA Alarm
Settings for calibration cylinder
Current date
Last calibration date (optional)
CAL due date (optional). If the calibration due date is activated, the
message "CAL DUE; X DAYS" appears on the instrument display.
X = the number of days until a calibration is due, user selectable for
1 to 180 days.
If the number of days until calibration is due reaches 0, an alert occurs
and "CAL DUE, NOW" displays.
Press the
instrument warm-up period.
Sensor warm-up period
Fresh Air Setup option (optional).
Refer to flowchart in Appendix, Section 7.1
US
button.
button to clear the alert and continue with the
ALTAIR 5 - Operating Manual
OPERATION
25