Push the threaded connection (2) over the condenser flange.
Push the spring (3) over the condenser flange.
Screw the G6 (6) gland on the drive head (Figure 4-10 (4).
Insert the G6 intermediate part (5) with seal G6 (4) into the gland G6 (6).
Screw the center piece (1) with the screw (2) on gland G6 (6).
Insert the condenser in the upper cut on the center piece (1).
4.5.4

Mount the vertical condenser (G3-G6) in the condenser bracket

This step is omitted for horizontal condensers (G1).
The support rod is installed.
Figure 4-12: Condenser bracket (Glassware set G3 and G6)
1
Support rod
2
Condenser clamp
Version glassware set G3 and G6:
1.
Push the boss head clamp (3) onto the support rod (1) and position with the clamp-
ing screw.
2.
Place the condenser clamp (2) around the condenser and attach on the support rod
by using the clamping screw.
Position the support rod with Allen screws (also refer to chapter 4.5.1).
Version glassware set G5 (depending on coolant):
Completely place the cold trap bracket into the lateral hole of the boss head
clamp (2).
Lock with the clamping screw.
Place the cold trap bracket on the liner cold trap jacket.
Position the liner cold trap jacket with the tension band on the cold trap bracket.
Position the support rod (1) with an Allen screw (also refer to chapter 4.5.1).
Hei-VAP Precision
1
2
3
3
Boss head clamp
Set-up and Start-up
Condenser mount-
ing
25
EN