EN
Operation of the Hei-VAP Precision
6.20.3
Auto
Menu entry
Heating bath
Rotation
Pump Type
Timer
Accept
Softkey 1
Softkey 4
6.20.4
Auto
Menu entry
Heating bath
Rotation
Distillation
rate (dT)
Hysteresis
Multiple sol-
vent
Final pressure
Timer
Accept
Softkey 1
Softkey 4
62
mode
easy
Value range
20 °C and the max. temperature of the heating bath (to be assigned in Setup), however, no
more than 210 °C.
10 and 280 rpm in 1 rpm increments; default value: 100 rpm
"vario control" and "vario tec" ; default value: "variocontrol"
Blank out by using vacuum-valve
1 to 1440 min; in 1 min; default value: 0 min; 0=off
Store Favorite
Esc
Mode
accurate
Value range / status
20 °C and the max. temperature of the heating bath (to be assigned in Setup), however, no
more than 210 °C.
10 and 280 rpm in 1 rpm increments; default value: 100 rpm
1 to 10
1-3 "low"
4-6 "medium"
7-10 "high"
Hysteresis value 1 to 50 mbar; default value: 10 mbar
yes/no; default value: No
1 mbar ... Safety pressure in 1 mbar; default value: 1000 mbar
1 to 1440 min; in 1 min; default value: 0 min; 0=off
Store Favorite
Esc
Hei-VAP Precision