Connecting A Contact Thermometer In Accordance With Din 12878 (Not Hei-Mix D) - Heidolph MR Hei-Tec Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

5
Connecting a contact thermometer in accordance with DIN 12878
(not Hei-Mix D)
Contact thermometers permit the temperature of the medium to be controlled in addition to
that of the hotplate.
Note:
We recommend the use of a safety contact thermometer when using the
instrument with a contact thermometer.
Pin assignment for safety contact
thermometer in accordance with
DIN 12878
Connect a 4-pin diode plug conforming to
DIN 41524 to the contact thermometer.
Refer to the drawing for the pin assignment.
The following accessories are available for use with a contact
thermometer or a Hei-Con:
X: Holder for thermometer or temperature sensor
(order no.:509-67000-00)
Y: Holder rod (order no.:509-81000-00)
Z: Hei-Con (order no.:509-88000-00) or
Hei-Con-G (order no.:509-88100-00)
Pin assignment for contact thermometer
without safety features
When using a contact thermometer without
glass breakage safety features, bridge pins
5 and 3.
Connect the
contact
thermometer to the
socket (I) on the
rear panel of the
instrument.
Remove the
jumper plug (J).
23
E

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Mr hei-mix d

Table des Matières