Error Code
Temperature Error
E9
Battery Error
E10
Tubing Disconnect Error
E12
A
Pressure Transducer Error
E13
Low Battery Error
Section V - Service and Maintenance
This service manual is intended for use as a guide to technically qualified personnel when evaluating System
malfunctions. It is not to be construed as authorization to perform warranty repairs. Unauthorized service will void the
warranty.
Introduction
WARNING: Do not service while the device is in use with a patient.
The Kendall SCD™ 700 series controller contains no user serviceable parts. User maintenance is covered in the sections
that follow. All other maintenance must be performed by technically qualified service personnel.
EN-17
Error Type
Manual reset required
E9
Service Required
User Resettable
B
Service Required
Recharge Battery
Description
If the internal case
temperature of the
controller drops below
5°C (41°F) or exceeds 55°C
(131°F).
Safe battery operation of
the controller can not be
ensured.
Pressure measured in
the inflatable garment is
below 10 mmHg for 10
consecutive cycles or no
garments are detected
during startup.
The system could not sense
a pressure rise of more
than 5 mmHg during an
inflation cycle or during
start up.
There is less than 15
minutes of battery charge
remaining. The pump and
valves will continue to
operate for as long as there
is enough power.
Troubleshooting
High temperature: Make
sure the controller is not
covered by bedding and
that the fan port, located
near the power cord is not
obstructed.
Low Temperature: Allow
the system to warm to
room temperature.
Service Technician
Only: Ensure that an
unauthorized battery
pack replacement has not
been made. Replace pack
or return to Covidien for
service.
Check for disconnected
tube sets or garments and
reconnect.
Service Technician Only:
Check the transducer tube
inside the controller and
ensure it is neither kinked
or disconnected.
Plug the controller into an
AC power outlet.
Kendall SCD™ 700 Series