Section III - Battery Operation

The Kendall SCD™ 700 series compression system is designed to operate normally on AC line power or DC battery
power without interruption. There are three Battery Status Indicator LED's used to represent the charge level of the
battery. Once the controller is powered on, it may take the system a few seconds to establish communication with the
battery and display the charge level. The battery Indicator shown below is located in the upper right hand corner of
the user interface. See FIGURE 10.
Warning: If the ground integrity of the mains power cable is in question, the device should be operated
on battery power until the ground integrity can be insured.

Unit plugged in and Powered On (Charging)

Battery State
100% charge
67-99% charge
34-66% charge
0-33% charge

Unit not plugged in and Powered On (Operating on Battery)

Battery State
67-100% charge
34-66% charge
< 34% charge
15-40 minutes left*
< 15 minutes left*

Unit Powered Off (charging when plugged in)

Battery State
0 -100% charge
Kendall SCD™ 700 Series
1
2
Battery Status Indicators
FIGURE 10
Battery Status 1
Green
Green
Green
Green (Pulsing)
Battery Status 1
Green
Green
Green
Amber (Flashing)
Red (Flashing)
Battery Status 1
Off
3
Battery Status 2
Green
Green
Green (Pulsing)
Off
Battery Status 2
Green
Green
Off
Off
Off
Battery Status 2
Off
Battery Status 3
Green
Green (Pulsing)
Off
Off
Battery Status 3
Green
Off
Off
Off
Off
Battery Status 3
Off
EN-12