Informational Messages
A
P
lArm
riority
Battery
Info
Discharging
Stopped Due
to
Temperature
Battery Not
Info
Charging
Due to
Temperature
Battery Not
Info
Charging
Check
Info
External
Battery
A
V
udible
isuAl
i
ndicAtor
And
•
"Batt Discharge Stopped -
Temp. " message
•
"Batt Not Charging -
Temp. " message
•
"Detach Battery Not
Charging" or "Internal
Battery Not Charging"
message
•
"Check External Battery"
message
(A
d
lArm
eVice
b
A
utton
ction
d
)
isPlAy
Operates
Operates
Operates
Operates
u
/c
ser
liniciAn
User: Press Alarm Indicator/Audio Pause
button to silence alarm. Or, press Left (Reset)
button to reset alarm. Move device to a cooler
location. Make sure device is not close to a
heat source. Make sure cooling vents are not
blocked. If alarm is detected for internal
battery and continues, place patient on
alternate source of ventilation and contact
your health care professional.
Clinician: If alarm is detected on the internal
battery and continues after the measures above
have been taken, place patient on alternate
source of ventilation and have device serviced.
User: Press Alarm Indicator/Audio Pause
button to silence alarm. Or, press Left (Reset)
button to reset alarm. Make sure device is
not close to a heat source. Make sure cooling
vents are not blocked. Move device to a cooler
location. If device is too cold, allow it to warm
up. If alarm continues, place patient on alternate
source of ventilation and contact your health
care professional.
Clinician: If alarm continues after the measures
above have been taken, place patient on alternate
source of ventilation and have device serviced
User: Press Alarm Indicator/Audio Pause
button to silence alarm. Or, press Left
(Reset) button to reset alarm. If condition
continues for internal battery, place
patient on alternate source of ventilation
and contact your health care professional.
Clinician: If condition continues for internal
battery, place patient on alternate source of
ventilation and have device serviced.
User: Press Alarm Indicator/Audio Pause
button to silence alarm. Or, press Left
(Reset) button to reset alarm. Check
connection to external battery. Replace
external battery with another external
battery, if available. If alarm continues, place
patient on alternate source of ventilation and
contact your health care professional.
Clinician: If the alarm continues after the
measures above have been taken, place
patient on alternate source of ventilation and
have device serviced.