Priority Alarm Sound Level Range Values
The following information is in addition to the Technical Specifications in Chapter 11 of the Trilogy manual.
The A-weighted alarm
sound power level range
The A-weighted alarm
sound pressure level range
Note: Range values determined per IEC 60601-1-8.
Pressure
The following specifications replace "Maximum Limited Pressure" under Pressure in Chapter 11 of the Trilogy manual.
Maximum Limited Pressure:
Control Accuracy
The following information replaces "Pressure Support" and related notations under the Control Accuracy Table in Chapter
11 of the Trilogy manual.
P
ArAmeter
Pressure Support
Note: Tolerances listed in the Control Accuracy Table here and in the Trilogy manual have a measurement
uncertainty of 3.60%.
3
The difference between the Inspiratory Pressure and the Expiratory Pressure must never be more than 40 cm H
Pressure Support and PEEP not to exceed 50 cm H
4
Measured Patient Parameter
The following section replaces the table and information under the Measured Patient Parameter Table in Chapter 11 of the
Trilogy manual.
P
ArAmeter
Vte / Vti
Minute Ventilation
Estimated Leak Rate
Respiratory Rate
Peak Inspiratory Flow
Peak Inspiratory Pressure
Mean Airway Pressure
% Patient Triggered Breaths
I:E Ratio
Note: Tolerances listed in the Measured Patient Parameter Table have a measurement uncertainty of 3.60%.
All flows and volumes are measured at BTPS conditions.
Medium Priority Alarm: 64.9 dB(A) to 92.9 dB(A) with an uncertainty of 2 dB(A)
High Priority Alarm: 65.4 dB(A) to 92.2 dB(A) with an uncertainty of 2 dB(A)
Medium Priority Alarm: 57.0 dB(A) to 85.0 dB(A) with an uncertainty of 2 dB(A)
High Priority Alarm: 57.3 dB(A) to 84.3 dB(A) with an uncertainty of 2 dB(A)
80 cm H
r
Ange
0 to 40 cm H
O
2
O.
2
r
Ange
0 to 50 ml
> 50 ml to 2000 ml
0 to 99 l/min
0 to 200 l/min
0 to 80 BPM
0 to 200 l/min
0 to 99 cm H
O
2
0 to 99 cm H
O
2
0 to 100%
9.9 - 1: 1 - 9.9
O (Volume Mode)
2
Greater of 2 cm H
3
Greater of 15 ml or 15% of reading
+/- (4.0 ml + 15% of the actual delivered volume)
Calculation based on measured Vte or Vti and Respiratory Rate
N/A
Greater of 1 BPM or 10% of reading
3 l/min plus 15% of reading
+ /- (2 cm H
O + 4 % of the actual reading)
2
+ /- (2 cm H
O + 4 % of the actual reading)
2
N/A
Calculation based on Inspiratory Time and Expiratory Time
A
ccurAcy
O or 8% of setting
2
A
ccurAcy
4
O.
2