INSTRUCTIONS FOR USE

Bottle with distilled water

8.3
The bottle with distilled water is located at the bottom of control panel. Distilled water from the bottle is used for the
syringe, micromotor, turbine, scaler on the dentist panel.
Replenishing distilled water:
turn the unit off
unscrew the bottle
replenish distilled water in the bottle
screw the bottle firmly to avoid compressed air leakage during the work
turn the unit on
check the bottle for air leakage
In the event that distilled water has been spent and there is air in the water distribution, it is recommended to deaerate
the water paths of instruments which use water by letting water flow until there are no bubbles in the water from
instruments.
Warnings
When replenishing distilled water, care must be taken, not to allow foreign substances penetrate into
the water or to cause changes in its quality or composition.
Distilled water for medical purposes must be used with maximal electrical conductivity of the water
up to 2000 µS/cm.
Do not use de-mineralized water for industrial purposes.
The manufacturer recommends to replace the bottle once a year.

Instruments cooling - setting of the water amount

8.4
To adjust water amount for instruments cooling, use solution which is applied on your unit:
Manual adjusting version
The amount of cooling water for all instruments, except for the syringe, can be adjusted by using of the needle valve
located on the bottom side of the control panel. When the valve is turned fully clockwise, the water supply stops.
Electronic adjusting version
If the dental unit is fitted with a proportional valve, the amount of the cooling water can be adjusted from the touch panel
for each instrument (except for the syringe). See illustration below.