C3
Fragile solids
•
Not wrapped
Rubber soids
•
121°C
Turbines if indicated by
the producer
•
Fragile hole instruments
•
C4 Wrapped
Wrapped rubber
•
121°C
Fragile solids
•
Wrapped and not
wrapeed turbines
•
Fragile hole instruments
•
C5
Not wrapped metal
SPECIAL
solids
134°C
It is compulsory to wrap the devices aimed at an invasive use (for instance; Surgical instruments,
etc.) to ensure a higher sterility.
The data of this table are indicative; the choice of the sterilisation cycle must be based on the data
supplied by the producer of the object you want to sterilise.
PAR. 9
TURN ON THE DEVICE PRESSING THE MAIN SWITCH.
Connect the suitable supplied hose to the loading plug and insert the other end (the one with the filter) inside
the demineralised or distilled water tank.
NOTE: The use of feedwater with contaminants at levels exceeding those given in the table above
can greatly shorten the working life of the device, with damage at the components of the autoclave
and then invalidate the manufacturer's warranty.
Pressing the button "16A", the pump is automatically turned on. By sucking, the pump fills the tank.
The pump supplies water for 120 seconds; after reaching the maximum level, it is turned off.
In case during these 120 seconds the pump does not reach the maximum level, the operator must press the
button "16A" up to switching off the pump in automatic mode (for the maximum reached level).
The autoclave does not work in case the tank level in on MIN. as signalled by the LED.
The attempt to start the autoclave with MIN. level state is visualised with the message "ER2" on the display
"TIME/MESS".
TABLE OF QUALITY LEVEL IN ACCORDING TO DIRECTIVE DIN EN 285
CEN STANDARD DIN EN 285
Evaporation residue
Silicium oxide, SiO2
Iron
Cadmium
Lead
Rest of heavy metals except iron, cadmium, lead
Chloride (Cl')
Phosphate (P20s)
Conductivity (at 20°C)
18 minutes
21 minutes
6 minutes
FILLING OF WATER TANK
21
17 minutes
2.8 Kg
Whole full load
17 minutes
2.8 Kg
Whole full load
2 minutes
2.8 Kg
Whole full load
10
1
0.2
0.005
0.05
0.1
2
0.5
15
(1.1÷1.2)
bar
(1.1÷1.2)
bar
(2.1÷2.2)
bar
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
s/cm
DNVT008 Rev_5