INSTRUCTIONS FOR USE
Button to correct the patient position
Press the button briefly to display the screen to control the basic movements of the chair and then the
patient position can be corrected. To return, press the button ESC or activate the instrument. It is still
possible to correct patient position (while instrument is activated) with the basic moves of the chair via foot
controller.
When a parameter is changed, while the instrument is taken out of its holder, without subsequent pressing the button
, the changed parameter will be saved only temporarily. The change will remain saved until the instrument
program has been changed (Prog 1 – 9). The change will be maintained ever after the dental unit has been turned on
/off. To cancel the temporary setting, press the button
be over-written with the data from the memory for the said program. To save the change of the parameter permanently
into the respective program, press the button
To activate the turbine, take it and move the foot controller lever to the right and/or the press the pedal of the combined
foot controller. When moved to the extreme position, the turbine has the ouput displayed. To stop the activity, move the
lever (pedal) of the foot controller to the starting position. Following the termination tool automatically goes blow through
for 0.5 seconds. CHIPBLOWER function is recommended to use after the activity has been finished. The turbine output
can be set by means of the buttons + / - within the range of 0-100% with the instrument taken in rest or in operation
and the lever /pedal of the foot controller is in maximum position. When working with the instrument, the output can be
adjusted continuously by pedal or by the lever of the foot controller in the analogue mode from 0 up to maximum set
value.
After the operation time of the turbine has reached the lubrication interval, the instrument lubrication alarm
displayed and audible signal sounds. This will be repeated each time the instrument is used until the information about
lubrication need is cancelled by pressing the button OK in the alarm screen.
Assign function to the left / right side button of the foot controller at instrument control
Pressing of left/right side icon of the button opens „pop-up menu" with the offer of functions. Select desired
function by pressing the appropriate button. Press „ESC" button if you want to cancel. You can setup both:
LEFT/RIGHT side buttons of the foot controller. You can choose only those functions, which are available
for the current tool. The others are disabled.
Chipblower function
Cooling of instrument ON/OFF
Changing of the direction of motor rotation
Instrument illumination ON/OFF
Main light ON/OFF
Bell
Cup filling
Bowl flushing
No function chosen
Switching the program of the instrument
Switch color of the instrument light (white/blue) (applies to motors: DENSIM DX BLUE, DENSIM DX
PRO BLUE)
UM_EN_TYGI350B_2018-09_ver1.0
and at the same time select the current program. Data will
.
after the change
TYGI 350 B