INSTRUCTIONS FOR USE
List of possible error conditions
Failure
Receiver not working - LED 1 is
not shining
The receiver works –
LED1 is lit, but the receiver
does not accept commands
from the foot controller
At the maximum of pedal lever,
the performance of the
instruments does not reach the
maximum
When the pedal is in the starting
position the instruments
continue to rotate
The pedal does not return to its
base position
Button for control of the chair
movement remains still pressed
Technical data:
Transmitter
Supply voltage
– via cable
– battery
Frequency
Battery life on a single charge
Transmitter reach
The transmitting power
Mudulation type
Receiver
Supply voltage
Frequency
UM_EN_TYGI350B_2018-09_ver1.0
Possible cause
The receiver is without
power
The receiver is
defective
Discharged battery of
the foot controller
The receiver, or the
transmitter is defective
Transmission barriers
Loose / damaged
potentiometer
Loose switch for start
Оut of position or
broken spring of pedal
Damaged foam spring
button
10-36 V DC (rated voltage 24 V)
2,4 V DC 2xAA NiMH 1200 mAh (GP Recyko+ or sanyo eneloop)
2400-2464 MHz
2 months
minimum 2 m
-18 dBm according to SW settings till 4 dBm
GFSK
24 V DC
2400-2464 MHz
Solution
Check if the dental unit is turned on
Check if receiver is properly connected with the dental
Connect the foot controller and the unit via cable and
contact an authorized service or reseller
Through the battery indicator on the dentist's display,
check the battery status
Charge the battery the foot of the driver
Contact an authorized service or reseller
Provide direct visibility between transmitter and receiver
Contact an authorized service or reseller
Contact an authorized service or reseller
Contact an authorized service or reseller
Contact an authorized service or reseller
TYGI 350 B
unit
40 / 49