Inlet Screen Installation

Inlet Screen Installation

The inlet screens mod. 9699304 and 9699316 prevent the blades of
the pump from being damaged by debris greater than 0.7 mm
diameter.
The inlet screen, however, does reduce the pumping speed by about
10%.
The inlet screen is fitted in the upper part of the pump, as shown in
the figure.
278/304
Turbo-V 1K-G User Manual / 87-901-025-01 (A)