Inlet screen installation

Figure 15
The inlet screen mod. 9699302 (ISO100-CFF6"), prevents the blades
of the pump from being damaged by debris. The inlet screen,
however, reduces the pumping speed as followed:
9699302
The inlet screen is fitted in the upper part of the pump, as shown in
the figure.
TwisTorr 404 FS User Manual / 87-901-048-01 (A.00)
Reduction in pumping speed (%)
Ar
N
2
20%
18%
Technical Information

Inlet screen installation

He
7%
17
H
2
5%
291/326