... the display
– The ambient temperature of the scale is too high or too low. Place the scale in an
ambient temperature between +10 °C and +40 °C. Wait about 15 minutes for the
scale to adapt to the ambient temperature and then weigh again.
E
... the display
and a number appear?
– Switch off the scale with the Start key and start the scale again. The scale will then
work normally again.
If this is not the case, disconnect the power supply by briefly removing the batteries. If
this measure is equally unsuccessful, inform the Service department.
... the weighing range won't switch?
– Take the load off the scale if necessary.
– Ensure that no special function is active.
9. Servicing/recalibration
Have the scale recalibrated by authorised persons in accordance with national legal re-
quirements. The year of initial calibration can be found behind the CE symbol on the rating
plate above the number of the appointed office, 0109 (Hessen Weights & Measures Of-
fice). Recalibration is essential if one or more security marks are damaged or the calibra-
tion counter does not match the number on the applicable calibration counter mark (cf.
page 24).
We recommend having your scale serviced by your local service agent before having it
recalibrated. seca Customer Service will be pleased to give you further assistance.
10. Technical data
Dimensions scale
width:
height:
depth:
Dimensions weighing platform
width:
height:
depth:
Weight
Temperature range
Calibration as per Directive
90/384/EEC
Medical product as per Directive
93/42/EEC
Colour
Maximum load
Weighing range 1
Weighing range 2
Model 675
appears?
890 mm / 35 ins
65
965 mm / 38 ins
800 mm / 31.5 ins
46
965 mm / 38 ins
approx. 25.5 kg
+10 °C to +40 °C
approved for medical use, Class III
Class I
silver grey/black
200 kg
300 kg
+10
+0.4
mm / 2.6
ins
+10
+0.4
mm / 1.8
ins
GB
31