Troubleshooting Guide - Medtronic Balance Affinity CP BBAP40 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 130
Check that all electrical connections are secure.
Test control module power and display.
Check date and integrity of sterile blood pump (and disposable probe if applicable) package(s).
If using the adapter, check the Use-by Date on the device label.
Check that flow transducer/sensor/disposable probe are sized properly.
Assemble perfusion circuit in a sterile manner.
Allow sufficient tubing length for standby drive unit or handcrank.
Connect flow transducer/sensor/disposable probe to circuit in correct location and flow direction.
2. PRIME PUMP AND CIRCUIT
Check circuit to ensure no kinks or occlusions are present.
CO
flush pump and circuit, if indicated; turn off CO
2
Gravity prime and debubble pump and perfusion circuit.
Check pump for leaks, irregular motion, and noise.
Check circuit for visible air.
Check that all tubing connections are secure.
Clamp pump outlet tubing completely.
Clamp venous return tubing completely.
3. OPERATING PARAMETERS
Perform the following, per the appropriate compatible system operator's manual:
Calibrate transducers/sensors according to the manufacturer's instructions.
Set and verify low/high flow alarms and other available safety systems.
4. EMERGENCY BACKUP EQUIPMENT
Ensure that the following equipment is available:
Backup power
Appropriate handcrank and drive unit from a compatible system
Spare Affinity CP centrifugal blood pump and disposable probe
If using the adapter, have a spare Affinity CP adapter available
Appropriate controller from a compatible system
5. PERFUSION
Achieve minimum pump flow prior to unclamping lines.
Monitor control module for messages and alarms.
Monitor perfusion circuit for visible air and kinks in the lines.
Maintain minimum pump flow prior to clamping lines.
6. CLEANUP
Turn off power.
Properly discard disposable components.
Clean motor and flow transducer/sensor.
If using the adapter, clean it according to the instructions in the Storage and Cleaning section.
7. CHECK EQUIPMENT
Inspect and verify that equipment is operational.
Maintain indicated preventive maintenance schedule.
Ensure batteries are charged to full capacity.

9. Troubleshooting Guide

This section covers 3 situations that may occur when using a centrifugal blood pump. Follow the clinical protocol specified by the
hospital and consider the following troubleshooting suggestions.
18
Instructions for Use
English
.
2

Publicité

Table des Matières
loading

Table des Matières