Straub Medical AG
®
Aspirex
S Catheter
WARNING: Before using the Straub Mechanical Thrombectomy System and its components, the user must be entirely familiar with the user
manuals of the Straub Medical Drive System and Straub rotational catheters.
Carefully read all instructions prior to use. Observe all warnings and precautions noted throughout these instructions. Failure to observe warnings
and precautions may result in complications and may lead to serious health consequences for the patient or to the patient's death.
Any recommendations within these instructions are designed to serve as a general guideline only. They are not intended to supersede
institutional protocols or professional clinical judgment concerning patient care.
Order number (REF), product name and materials included in sets:
REF
REF
Name
®
80209
Aspirex
S 6F 110cm
®
80203
Aspirex
S 6F 135cm
®
80210
Aspirex
S 8F 85cm
®
80204
Aspirex
S 8F 110cm
®
80211
Aspirex
S 10F 110cm
80233
GW 0.018" 4/220cm
80234
GW 0.018" 4/270cm
80235
GW 0.018" 4/320cm
80010
GW 0.032" 10/260cm
10170
Collecting bag, 2000ml
80068
Drape, sterile for motor or
footswitch
11121
Instructions for use
®
Aspirex
S
Technical details and specifications
Designed for the treatment of
these vessel diameters:
Catheter external diameter:
Rotation:
Maximum aspiration capacity
(maximum blood loss):
Power transmission:
Indication for use:
®
Aspirex
S catheters in combination with the Straub Medical Drive System (REF SRS-Set) are licensed for the percutaneous transluminal
removal of fresh thrombotic or thromboembolic material from blood vessels outside the cardiopulmonary, coronary and cerebral
circulations.
Indicated for:
Materials not included in delivery (needed for use):
Introducer sheath
Heparinised isotonic saline
Storage and handling:
The products must be stored in their original packaging in a dry, dark place. The suitable temperature range for storage is 10-40°C (50-104°F).
11121 A1
80226
80227
®
SET Aspirex
S
SET Aspirex
6F 110cm
6F 135cm
1
-
-
-
-
-
1
-
-
1
2
1
3-5 mm
2 mm
60,000 r.p.m.
45 m/min
Native blood vessels or vessels fitted with stents, stent grafts or native or artificial bypasses outside the
cardiopulmonary, coronary and cerebral circulations.
Use an introducer sheath fitting the patient's anatomy as well as the catheter's dimension.
WARNING: Only use sheaths that are highly resistant to kinking. A kink in the introducer sheath
occurring during the Aspirex
even break off, which can have serious health consequences for the patient.
CAUTION: The internal lumen of the introducer sheath must at least correspond to the external
diameter of the catheter. Using an introducer sheath with a larger internal lumen can simplify handling
of the catheter.
CAUTION: Follow the operating instructions for the sheath selected.
For filling the catheter lumen with heparinised isotonic saline before use and to flush the catheter when the
procedure is intermitted.
EN
Instruction For Use
80229
®
®
S
SET Aspirex
S
8F 85cm
-
-
1
-
-
1
-
-
-
-
-
1
-
-
1
-
-
-
1
1
2
2
1
1
Magnetic clutch
®
S procedure can cause damage and parts of the Aspirex
© Straub Medical AG
0408
80230
®
SET Aspirex
S
8F 110cm
-
-
-
1
-
-
1
-
-
1
2
1
5-8 mm
2 ⅔ mm
40,000 r.p.m.
75 ml/min
80232
®
SET Aspirex
S
10F 110cm
-
-
-
-
1
-
-
-
1
1
2
1
8-12 mm
3 ⅓ mm
40,000 r.p.m.
180 ml/min
®
S catheter may
3 / 76