9 Replace canister
WARNINGS
The canister is replaced on the basis of a visual check or according to the instructions on
the display (warning signal). Wear gloves for all operations.
The warning «canister full» is triggered only once. After cancelling the warning with
the selection buttons, the warning is deactivated and a symbol [
mode. After deactivation of the warning, the user is responsible for checking the canister
filling level. Please refer to chapter "13 – Troubleshooting and alarms" for a more detailed
explanation of the canister full warning.
30 |
Replace canister
1. Prepare sterile canister (with/without solidifier).
2. Close patient tubing with the tube clamp.
3.
4.
5.
6.
Canister change
] flashes in the run
3.1 Switch to Standby mode
by pressing «Standby» for
longer than 3 seconds.
4.1 Release the canister by
pressing the release button.
4.2 Remove canister.
5.1 Unpack the canister.
5.2 Position the openings at
the top and attach the
bottom part of the canister
to Thopaz.
6.1 Push the canister into
Thopaz until you hear
a click.
Safety related tip
Using a 2 l canister
refer to chapter 8.