WARNINGS
– Consult this manual prior to use.
– Data transfer via USB is not allowed during therapy.
– Before cleaning the device, pull the plug out of the fixed mains socket.
– Do not use Thopaz in MRT (Magnetic Resonance Tomography).
– Do not dry Thopaz with microwaves.
– The device is not suitable for use while bathing, showering or in a hazardous explosive
environment.
– The pressure range to be set must be determined by a physician in accordance with
the age and weight of the patient.
– The secretions gathered must be monitored more regularly or frequently in accordance
with the age and weight of the patient.
– To ensure Thopaz can reduce the pressure, it is essential to wait 30 seconds between
taking the sample from the sample port and unclamping the patient catheter (drain).
– Do not use Thopaz if the drainage therapy indicates a pressure greater than Thopaz
max. pressure range of -10 kPa.
– Do not use Thopaz if the drainage therapy indicates a flow rate greater than Thopaz
max. flow capacity of 5 l/min.
– Do not use Thopaz if the drainage therapy indicates no pressure should be applied to
the patient.
– If a persistent air leak is shown, check the system is assembled correctly before taking
further corrective actions. Ensure the system is air tight by clamping the catheter and
observing the air leak decrease to zero.
– The catheter/connector interface is a location where clotting may occur.
We recommend regular monitoring of this interface and appropriate removal
procedure in case of occurance.
– Do not connect bilateral thoracic drains to one Thopaz unit. In such cases, the use of
two Thopaz units is recommended.
– The use of Thopaz for any other indication than intended is neither desired nor allowed.
– The Thopaz system is not designed for retransfusion.
– This product contains lithium-ion batteries which bear risk of fire, explosion and burns.
Do not dissassemble, crash, heat above 100 °C (212 °F), incinerate or dispose of in fire.
– For pediatric patients adapt pressure settings according to hospital guidelines.
– Do not touch the patient and the docking contacts simultaneously.