Change canister and tubing
XV

Change canister

- on the basis of a visual check
- according to the instructions in the display / alarm signal
4
5
8
1. Provide sterile canister.
2. Clamp patient tubing.
3. Switch off pressure with the "Standby" button (press for > 3 sec.).
4. Release and remove canister.
5. Unpack new canister, position and click into Invia Liberty.
6. "On". Pressure is built up.
7. Unclamp tubing.
8. Seal used canister and dispose of it in accordance with
disosal procedures of the facility.
17
EN