13 Troubleshooting and alarms

Pump not running

Check if:
– the mains voltage is correct and if the mains plug is inserted correctly into the fixed
mains socket.
– Thopaz is switched on.

Description

Thopaz distinguishes between warnings, alarms and internal errors. If Thopaz detects any of
these situations, an acoustic warning signal sounds and a description of the problem appears
on the display. By pressing the two selection buttons simultaneously, the acoustic alarm is
suppressed for 60 seconds and the troubleshooting instructions appear on the display.
Warning
Pressure is maintained.
beep
Problem described on display
Example:
Troubleshooting instructions
are displayed
36 |
Troubleshooting and alarms
Alarm
Pressure is interrupted.
beep
to acknowledge an acoustic alarm
Press simultaneously, then follow the instructions on
the display
Error number
Internal error
Problem cannot be
rectified by the user.
The error number is
beep
displayed.