Carry out functional
check
Safety related tip
The acoustic alarm function «Leak in system – 301» is deactivated for approximately
6 minutes after Thopaz switching on.
1.
2.
Flow decreases:
Thopaz is ready for use.
Flow does not decrease:
Leak in the system; check tubing connections. Check pres-
sure relief value for tight connection. Replace canister and/
or tubing if necessary.
3.
4.
Thopaz is in Standby mode and is ready for use with
the factory settings.
(To change factory settings please refer to chapter
11-Changing settings)
5.
Connect the Thopaz system to the patients catheter (drain)
according to hospital guideline. To start the therapy press «on».
Preparation for use / operating instructions
1.1 Seal off the conical con-
nector of the patient tub-
ing with your thumb
(through the packaging).
The tubing must not be
kinked otherwise the mea-
suring tube will be closed.
2.1 Switch on pressure by
pressing «on» – pressure
is built up.
2.2 Check flow value.
3.1 Switch to Standby mode
by pressing «Standby» for
longer than 3 seconds.
| 25