Cleaning; Maintenance And Service - Soehnle Professional 3710 Standard Série Mode D'emploi

Table des Matières

Publicité

not for use in environments where there is a risk of explosion. The mains-operated
display unit must be set up so that it can be easily disconnected from the mains power
(ensure that the socket is accessible). The display unit should be protected from
moisture.
This device has interference suppression in accordance with the related EC guideline
2014/30/EU. However, the display reading can be affected by extreme electrostatic or
electromagnetic interference, for example if a radio or mobile phone is used
immediately beside it. If affected in this way, the device should return to normal
function once the source of interference is removed. However, it may be necessary to
switch off and on again. If there is permanent electrostatic interference, please
contact your service partner.
This device is a measuring instrument. Drafts, vibrations, rapid temperature changes
and direct exposure to sunlight may affect the weight readings.
The scales conform to protection class IP42. Avoid high levels of air humidity, steam,
aggressive fluids and heavy soiling.
Improper installation will invalidate the warranty. The electrical connection conditions
must be consistent with the rating values printed on the power adapter. The
equipment is designed to be used indoors. Note the permitted ambient temperatures
for operation (see Technical Data). This equipment meets the requirements for
electromagnetic compatibility. Do not exceed the maximum values specified in the
standards. If used correctly, there is no danger of other equipment being affected by
this product. If you have any problems, please contact your Soehnle Professional
service partner.
Warning: This equipment must not be modified without the manufacturer's
permission. If medical equipment is modified, suitable tests and inspections must be
performed to ensure that it can still be used safely.

5.4 Cleaning

Before cleaning, disconnect the equipment from the power supply by pulling out the
mains plug. The equipment should be cleaned only with a moistened cloth. In no
circumstances should water be allowed to penetrate the equipment.

5.5 Maintenance and service

Metrological inspection was carried out at the initial verification when the product was
manufactured. Other regular metrological controls (recalibrations) are to be carried
out by the competent weights and measures authorities according to the relevant
national regulations. Repairs and replacement of the power cable must be carried out
only at a centre authorised by Soehnle Professional, and using original parts.
The device should be undergo preventative inspections regularly (6 monthly intervals).
The scales are to be loaded with a known weight and the value displayed compared
www.soehnle-professional.com
45