4. FAULTS – CAUSES AND REMEDIES
If the fault cannot be cleared, or other error messages appear, please contact your
service partner.
www.soehnle-professional.com
DISPLAY
When switched on, the scales are set
automatically to zero. If the scales are
outside of the designed tolerances the
display will show -0-.
Display not reset to 0 when switched
on.
Underload: only dashes appear at the
bottom of the display field.
Overload: only dashes appear at the
top of the display field. Maximum
weighing range is exceeded.
Zero setting not possible.
Weight has gone above or below zero
setting limit.
Taring not possible with the scales
unloaded, not yet stabilised, or
overloaded.
Printing not possible in the case of
underload or overload.
kg/lb switching blocked.
Battery discharged. About 30 minutes
operating time remaining.
Battery flat.
REMEDY
Unload the platform. Remove
any dirt if present. If the scales
do not display a zero after a few
seconds, please contact your
service partner.
Switch off the display. When
switched on, zero point is reset
automatically.
Unload the scales, or switch
them off and on again. The zero
point is reset automatically.
Check the surface under the
scales, and check that the
weighing platform is level. Test
the scales for a force shunt. If
the error persists after
readjustment, please contact
your service partner.
Stabilise the weight on the
scales or turn off or on with
unloaded scales.
Connect the equipment to the
power adapter to recharge.
Connect the equipment to the
power adapter to recharge.