8. Malfunctions - Causes and elimination

Read the following troubleshooting instructions before contacting service personnel.
Error
The scale cannot be switched
on when using the battery.
The scale cannot be switched
on when using the rechargeable
battery.
Displayed weight questionable
or scale does not reset to zero.
„STOP" is shown in the display.
„---" is shown in the display.
„LO Bat" is shown in the display. Rechargeable battery is weak.
Error list
Error
SAT
RTC
If the faults cannot be rectified or if other error messages are displayed, please contact your your
service partner.
Possible cause
Battery discharged
1. Battery discharged
2. Mains socket defective
3. Power supply defective
1. Weight display is falsified by
object
2. Weighing tray does not rest
correctly
3. Before weighing, the display did
not show „0.0 „lb/kg
4. Scale not set up stably
5. Scale no longer correctly
calibrated
Load exceeds the capacity of the
scale.
Scale is in underload condition.
Definition
A/D saturation
Reading error of the real time clock
Remedy
Connect the scale to the power
source.
1. Replace batteries
2. Use another mains socket
3. Replace power supply unit
1. Remove all objects / infant from
the weighing tray
2. Arrange weighing tray correctly
3. Remove the infant from the scale,
set the scale to zero and weigh
again
4. Place the scale on a stable surface
and repeat the weighing procedure
5. Check weight with object of known
weight
Remove excessive weight and obser-
ve maximum load values.
Check that the weighing tray is on
the base.
Recharge the battery according to
the instructions.
Check / Replace
Check charging cell
– CONSULT A TECHNICIAN
Battery of the real-time clock empty
or circuit board defective
– CONSULT A TECHNICIAN