4. FAULTS – CAUSES AND REMEDIES
 
 
 
 
If you are unable to correct the fault or if other error messages appear, please contact your service
partner.
 
www.soehnle-professional.com
DISPLAY
When switched on, the scales are set
automatically to zero. If the scales are
outside the specified tolerance range,
the display shows -0-.
Underload: only dashes appear at the
bottom of the display field.
Overload: only dashes appear at the top
of the display field. Maximum weighing
range is exceeded.
Part weight is too low.
Weight has gone above or below zero
setting limit.
Taring not possible with the scales
unloaded, not yet stabilised, or
overloaded.
Printing not possible in the case of
underload or overload.
kg/lb switching blocked.
 
REMEDY
Unload the platform. Remove any
dirt if present. If the scales do not
display a zero after a few
seconds, please contact your
service partner.
Switch the scales off and then on
again. The zero point is reset
automatically.
Remove part of the weighed
goods from the scales.
 
48