14.3

Setting options

t/min
Runtime . Adjustable from 1 - 99 min, in steps of 1 minute.
t/sec
Runtime . Adjustable from 1 - 59 s, in steps of 1 second.
Continuous run " --:-- ". Set the parameters t/min and t/sec to zero.
RPM
Speed
A number value from 200 rpm to the maximum speed of the rotor can be set.
Adjustable from 200 rpm to 10000 rpm in steps of 10, and starting from 10000 rpm to the maximum speed,
in steps of 100.
For the maximum rotor speed, see the chapter "Appendix, Rotors and accessories".
RCF
Relative centrifugal acceleration
A number value can be set which results in a speed between 200 rpm and the maximum rotor speed.
Adjustable up to 10000 in steps of 1, and starting from 10000, in steps of 10.
It is only possible to input the relative centrifugal acceleration (RCF) if the RCF display ( RCF ) is
activated.
The relative centrifugal acceleration (RCF) depends on the centrifuging radius (RAD). After entering
the RCF, check to make sure that the correct centrifuging radius has been set.
RAD/mm Centrifuging radius
Adjustable from 10 mm to 250 mm, in steps of 1 mm.
For the centrifuging radius, see the chapter "Appendix, Rotors and accessories".
It is only possible to input the centrifuging radius if the RCF display ( RCF ) is activated.
DEC
Brake stage. fast = short run-out time, slow = long run-out time.