Setting Options - Hettich EBA 280 Mode D'emploi

Masquer les pouces Voir aussi pour EBA 280:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

14.3

Setting options

PROG RCL
Program location of the program to be called
The program locations 1 to 9 and # can be called.
Runtime. Adjustable from 1 - 99 min, in steps of 1 minute.
t/min
Runtime. Adjustable from 1 - 59 s, in steps of 1 second.
t/sec
Continuous run "--:--". Set the parameters t/min and t/sec to zero.
Begins at
Counting the runtime.
The runtime begins counting immediately after the centrifugation run begins.
START
The runtime only begins counting after reaching the set speed.
Begins at
SPEED
This setting is only possible when the "Dual time" function is activated. See the
"Activating/deactivating the 'Dual time' function" chapter. This function is activated ex works.
RPM
Speed
A number value from 200 rpm to the maximum speed of the rotor can be set.
Settable in increments of 10.
For the maximum rotor speed, see the chapter "Appendix, Rotors and accessories".
RAD/mm
Centrifuging radius
Adjustable from 10 mm to 330 mm, in steps of 1 mm.
For the centrifuging radius, see the chapter "Appendix, Rotors and accessories".
It is only possible to input the centrifuging radius if the RCF display ( RCF ) is activated.
RCF
Relative centrifugal acceleration
A number value can be set which results in a speed between 200 rpm and the maximum rotor speed.
Settable in increments of 1.
It is only possible to input the relative centrifugal acceleration (RCF) if the RCF display ( RCF ) is
activated.
The relative centrifugal acceleration (RCF) depends on the centrifuging radius (RAD). The relative
centrifugal acceleration RCF) and the set centrifuging radius (RAD) are displayed.
Starting steps 1 - 9. Step 9 = shortest starting time, Step 1 = longest starting time.
Brake steps 0 - 9. Step 9 = shortest run-down time, Step 1 = long run-down time,
Step 0 = longest run-down time (brakeless run-down).
PROG STO
Program location where the program is stored.
9 different programs can be stored (program locations 1 to 9).
No programs should be saved at program location #.
The program location # serves as intermediate memory for changed centrifugation parameters.
The centrifugation parameters at program location # are overwritten every time a centrifugation
run is carried out with changed centrifugation parameters which were not saved to a program
location.