5.2. Specifications
Infant
Resuscitator volume
approx. 215 ml
Delivered volume one hand*
150 ml
Delivered volume two hands*
-
Dimensions
approx.
(length x diameter) w/o
190 x 71 mm
reservoir and accessory
Weight w/o reservoir
approx. 70 g
and accessory
4.0 kPa
Pressure-limiting valve**
(40 cmH₂O)
≤ 5 ml + 10 % of the
Dead space
delivered volume
max 0.1 kPa
Inspiratory resistance***
(1.0 cmH₂O)
at 5 l/min
max 0.2 kPa
Expiratory resistance***
(2.0 cmH₂O)
at 5 l/min
approx. 300 ml (bag)
Reservoir volume
approx. 100 ml (tube)
Patient connector
Expiration connector
(for PEEP valve attachment)
Manometer Port connector
Demand Valve connector
-
Pediatric
Adult
approx. 664 ml
approx. 1547 ml
450 ml
600 ml
-
1000 ml
approx.
approx.
223 x 99 mm
284 x 127 mm
approx. 145 g
approx. 220 g
4.0 kPa
4.0 kPa
(40 cmH₂O)
(40 cmH₂O)
≤ 5 ml + 10 % of the
≤ 5 ml + 10 % of the
delivered volume
delivered volume
max 0.5 kPa
max 0.5 kPa
(5.0 cmH₂O)
(5.0 cmH₂O)
at 50 l/min
at 50 l/min
max 0.27 kPa
max 0.27 kPa
(2.7 cmH₂O)
(2.7 cmH₂O)
at 50 l/min
at 50 l/min
approx. 2600 ml (bag)
approx. 2600 ml (bag)
Outside 22 mm male (ISO 5356-1)
Inside 15 mm female (ISO 5356-1)
30 mm male (ISO 5356-1)
Ø 4.2 +/- 0.1 mm
Inside 32 mm female (EN ISO 10651-4)
Infant
Forward and
backward leakage
M-Port
O₂ inlet connector
Operation temperature limits -18 °C to +50 °C (-0.4 °F to +122 °F), tested according to EN ISO 10651-4
Storage temperature limits
-40 °C to +60 °C (-40 °F to +140 °F), tested according to EN ISO 10651-4
Recommended long term storage in closed packaging at room temperature, away from sunlight.
* Tested according to EN ISO 10651-4.
** Higher delivery pressure can be obtained by overriding the pressure-limiting valve.
*** At general test conditions according to EN ISO 10651-4.
5.3. MRI Safety Information
The Ambu SPUR II Resuscitator
, and Ambu SPUR II Resuscitator with attached Ambu
1
Disposable PEEP 20 Valve
, and Ambu SPUR II Resuscitator with attached Ambu Disposable
2
PEEP 20 Valve and Ambu Disposable Pressure Manometer
therefore may be safely used in the MR environment (not inside the MR bore) under
the following conditions.
Static magnetic field of 7 Tesla and less, with
Maximum spatial field gradient of
– 10,000 G/cm (100 T/m)
1
– 16,000 G/cm (160 T/m)
2
Maximum force product of
– 450,000,000 G
/cm (450 T
/m)
2
2
1
– 721,000,000 G
/cm (721 T
/m)
2
2
2
Use inside the MR bore may influence MR image quality.
RF-induced heating and MR image artifacts have not been tested. Any metallic parts are
fully encapsulated and do not have any contact with the human body.
Pediatric
Adult
Not measurable
Connector compatible with EN ISO 80369-7
According to EN 13544-2
2
are MR Conditional, and
11