Télécharger Imprimer la page

Ambu SPUR II Mode D'emploi page 10

Masquer les pouces Voir aussi pour SPUR II:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 59
Measuring EtCO₂
For measuring of side stream EtCO₂; connect the gas-sampling line for the EtCO₂
measuring device to the M-Port of Ambu SPUR II Resuscitator. Connect the gas sampling
line connector by mounting and rotating it 1/4 turn clockwise.
Applying medication
Carefully observe patient response to the administered medication. Administration of
volumes of 1 ml fluid or above through the M-Port is comparable with administration
directly into an endotracheal tube. The M-Port has been tested with epinephrine,
lidocaine, and atropine.
Ambu SPUR II Resuscitator demand valve version
Ambu SPUR II Resuscitator demand valve version is available in adult and pediatric sizes
and can be used either with or without a demand valve and comes with an attachable
oxygen reservoir bag.
The inlet valve of the Ambu SPUR II Resuscitator connects to the demand valve via an adapter.
Attachment of the demand valve:
– Remove the oxygen reservoir unit from the Ambu SPUR II Resuscitator inlet valve, if attached.
– Attach the adapter to the demand valve system.
– Insert the demand valve adaptor in the Ambu SPUR II Resuscitator inlet valve.
NOTE: Use only part labeled "compression unit" with adapter and Demand valve. The part labeled
"Oxygen Reservoir Bag" is regarded as back-up if the demand valve fails.
Oxygen administration
Administer oxygen according to medical indication.
The figure
6
shows Calculated delivered oxygen percentages which can be obtained
with different ventilation volumes and frequencies at different gas flow rates. The oxygen
percentages can be seen in
Adult
6
6.1
10
5
, Pediatric
, Infant
.
6.2
6.3
Oxygen reservoir bag
7
The thin oxygen reservoir bag plastic foil cannot be detached from its point of fixation to
the resuscitator at any time.
Accessories
8
The Ambu SPUR II Resuscitator connectors follows ISO 5356-1 and EN 13544-2 making it
compliant to other hospital equipment. When applying external devices, make sure to test
for functionality and consult the instructions for use accompanying the external device.
Ambu products compatible with Ambu SPUR II Resuscitator are stated below:
Ambu® Disposable Face Mask
For further information please refer to the instruction for use of the Ambu Disposable
Face Mask.
Ambu® Disposable PEEP 20 valve
8.1 8.2
For further information please refer to the instruction for use of the Ambu Disposable PEEP
20 valve or refer to illustration
in this instruction for use. To fit the Ambu Disposable
8.1
PEEP 20 valve (if required) to the resuscitator, remove the splash guard.
Ambu® Disposable Pressure Manometer
For further information please refer to the instruction for use of the Ambu Disposable
Pressure Manometer.
4.4. After use
Used products must be disposed of according to local procedures.
5. Technical product specifications
5.1. Standards applied
The Ambu SPUR II resuscitator is conforming with the product specific standard EN ISO 10651-4.
8.2
8.3

Publicité

loading