WARNING
Risk of explosion!
If the CatEx sensor is exposed to hydrogen and low
temperatures (< -10 °C) for a long time (> 1 h), measured
values that are too low may be displayed. This also applies if
hydrogen is being measured but the measured gas set is not
hydrogen.
► The CatEx sensor should be set to the measured gas
"hydrogen"for regular and planned measurements of
hydrogen. Measuring tasks which last less than 30 min.
are not affected by this.
Increased hydrogen concentrations within the measuring
range of the DrägerSensor XXS H
may trigger false alarms by additively influencing Dräger
sensors XXS H
S and XXS CO, XXS H
2
LC, and negatively influencing DrägerSensor XXS O
XXS O
PR.
2
2
Conventions in this document
2.1
Meaning of the warning notes
The following warning notes are used in this document to
notify users of possible dangers. The meanings of the warning
notes are defined as follows:
Alert icon Signal word
WARNING
CAUTION
NOTICE
2.2
Trade marks
Brand
Brand owner
Dräger
®
X-am
Bluetooth SIG, Inc.
®
Bluetooth
The brands listed are only registered in certain countries and
not necessarily in the country in which this material is sold.
2.3
Abbreviations
Abbreviation
Explanation
A1
Pre-alarm
A2
Main alarm
Instructions for use
|
X-am
HC and CatEx sensor
2
S-LC and XXS CO-
2
2
Consequences in case of
nonobservance
Indicates a potentially hazard-
ous situation. If not avoided, it
could result in death or serious
injury.
Indicates a potentially hazard-
ous situation. If not avoided, it
could result in physical injury. It
may also be used to alert
against unsafe practices.
Indicates a potentially hazard-
ous situation. If not avoided, it
could result in damage to the
product or environment.
®
®
2800 / X-am
5800
Abbreviation
PID
STEL
TWA
3
Description
3.1
Product overview
and
1
2a
2
7
6
OK
Graphic A
1 Gas inlet
2 Alarm LED
2
D-Light
a
3 Horn
4
key
OK
5 Power pack
6
key
7 Display
Marking added-signal channels (X-am 5800 only):
Feature
Toxic Twins
CO H
compensation
2
Conventions in this document
Explanation
Photoionization detector
Short time exposure limit, threshold of an
exposition over a short period of time (gen-
erally 15 minutes).
Time weighted average, average shift val-
ues are generally limited to eight hours
exposure per day per workplace for 5 days
a week during a work lifetime. Observe the
national definition of the occupational
exposure limit.
8
2
9
3
10
4
12
5
11
8 IR interface
9 Fastening clip
1
Name plate
0
1
Charging contacts
1
1
Measured gas display
2
1
Measuring value display
3
1
Special symbols
4
1
Allen key
5
(2 mm; alternative: Torx
T8)
Shown in display
HCN+
CO+
|
en
13
14
15
21