Measurement system uncertainties
In accordance with ISO 80601-2-70:2020 and ISO 80601-2-79:2018 the measurement uncertainty of the manufacturer's test
equipment is:
For measures of flow:
For measures of volume:
For measures of static/dynamic pressure:
For measures of time:
In accordance with ISO 80601-2-74:2021 the measurement uncertainty of the manufacturer's test equipment is
For measures of humidification output:
Bluetooth
Technology used:
Connection types:
Frequency:
Max RF power output:
Operation range:
Cellular technology and regulatory compliance
Refer to the Cellular information guide in ResMed.com/downloads/devices.
The device should be installed and operated with minimum distance 15 mm (0.59") between the equipment and the user's body.
Humidifier
Maximum heater plate temperature:
Temperature cut-out (heater):
Maximum gas temperature (at mask)
1
The air flow for breathing produced by this therapy device can be higher than the temperature of the room. Under extreme
ambient temperature conditions (104ºF/40ºC) the device remains safe.
Humidifier performance
SlimLine/Standard tubing
Mask Pressure
Nominal RH output % at 72ºF (22ºC) ambient
cm H
O (hPa)
temperature
2
Setting 4
(default setting)
3
80%
4
80%
10
80%
20
80%
25
80%
Climate Control Auto - ClimateLineAir 11
Mask Pressure
Nominal RH output % at 72ºF (22ºC) ambient
cm H
O (hPa)
temperature
2
3
85%
4
85%
10
85%
20
85%
25
85%
1
AH - Absolute Humidity in mg/L
2
BTPS - Body Temperature Pressure Saturated
3
Humidifier performance meets ISO 80601-2-74:2021 performance > 10 mg/L BTPS tested at 59°F to 95°F (15°C to 35°C)
± 3.9 L/min
± 6 mL or 5% (whichever is greater)
± 0.15 cm H
± 6 ms
± 0.5 mg/L BTPS
Bluetooth Low Energy (BLE)
GATT
2400 to 2483.5 MHz
+4 dBm
10 m (Class 2)
154ºF (68ºC)
165ºF (74ºC)
1
≤ 106ºF (41ºC)
:
Setting 8
(maximum setting)
100%
100%
100%
100%
100%
O (± 0.15 hPa)
2
Nominal system output mg/L AH
Setting 4
(default setting)
≥6
≥6
≥6
≥6
≥6
Nominal system output mg/L AH
≥ 12
≥ 12
≥ 12
≥ 12
≥ 12
1
2
, BTPS
Setting 8
3
(maximum setting)
≥12
≥12
≥12
≥12
≥12
1
2
, BTPS