Problem/possible cause
My therapy data has not been sent to my care provider/prescription settings have not been applied to my device
Wireless coverage may be poor/The no wireless
connection icon
is displayed on the top right of the
screen.
Device may be in Airplane Mode.
Data transfer is not enabled for your device.
SmartStart is enabled, but the device does not automatically start when I breathe into the mask
Breath is not deep enough to trigger SmartStart
There is excessive leak
SmartStop is enabled, but does not automatically stop when I remove the mask.
Incompatible mask being used
Device Messages
Problem/possible cause
High leak detected. Check your humidifier or side cover.
Humidifier tub or side cover may not be inserted properly
High leak detected. Connect your tubing.
Air tubing may not be connected properly.
Mask may be fitted incorrectly.
Tubing blocked. Check your tubing.
Air tubing may be blocked
Read only card, remove, unlock and re-insert SD card
SD card switch may be in the lock (read-only) position
Solution
Make sure that the device is placed where there is coverage (ie, on
your bedside table, not in a drawer or on the floor).
The wireless signal strength icon
when all bars are displayed, and poor coverage when fewer bars are
displayed.
Turn off Airplane Mode.
Talk to your care provider about your settings.
To start therapy, take a deep breath in and out through the mask,
before breathing normally.
Press the Start Therapy/ Standby button located on top of the device.
Adjust the mask and headgear
Air tubing may not be connected properly. Connect firmly at both ends.
Only use equipment recommended by ResMed.
Contact ResMed or see ResMed.com for more information.
If you are using a conduit mask, SmartStop may not work.
Solution
Make sure the humidifier tub or side cover is correctly inserted
Make sure the air tubing is firmly connected at both ends.
Make sure the mask is fitted correctly. See the mask user guide for
fitting instructions or use the Mask Fit function to check the mask fit
and seal.
Check the air tubing and remove any blockages. Press the Start
therapy/standby button to restart the device
Move the switch on the SD card from the lock position
unlock position
and then re-insert it.