ResMed AirCurve 11 VAUTO Mode D'emploi page 29

Masquer les pouces Voir aussi pour AirCurve 11 VAUTO:

Publicité

Les langues disponibles

Les langues disponibles

Value
1
Pressure at mask
:
Mask pressure
Flow and flow derived values
Flow
Leak
Tidal volume
Respiratory rate
Minute ventilation
1
Results are expressed as STPD (Standard Temperature and Pressure, Dry). Use the following table to convert the STPD flow setting to BTPS
(Body Temperature and Pressure, Saturated) flow.
2
Accuracy may be reduced by the presence of leaks, supplemental oxygen, tidal volumes <100 mL or minute ventilation <3 L/min.
STPD to BTPS conversion
Altitude (m)
0
500
1000
1500
2000
2500
3000
Pressure accuracy as tested according to ISO 80601-2-79:2018
± (0.5 cm H
O (hPa) + 4% of the set pressure) cm H
2
Pressure accuracy - CPAP mode
Maximum static pressure variation at 10 cm H
Device with HumidAir 11 humidifier tub / side cover and air tubing: ±0.5 cm H
Maximum dynamic pressure variation according to ISO 80601-2-70:2015
Device with HumidAir 11 humidifier tub / side cover and air tubing:
Breath rate
Dynamic pressure variation (cm H
Maximum dynamic pressure variation according to ISO 80601-2-70:2020
Device with HumidAir 11 humidifier tub / side cover and air tubing:
Maximum error from set pressure (cm H
Note: Refer to the relevant measurement uncertainty from the Measurement system uncertainties table.
Pressure accuracy - Bi-level modes
Maximum dynamic pressure variation according to ISO 80601-2-70:2015 and ISO 80601-2-70:2020
Device with HumidAir 11 humidifier tub / side cover and air tubing:
Inspiration/Expiration mean error ± standard deviation (cm H2O [hPa]): 1±0.1
Note: Refer to the relevant measurement uncertainty from the Measurement system uncertainties table.
% of Inspiratory Phase for calculation: > 60
% of Expiratory Phase for calculation: > 66
Note: For each inspiratory and expiratory breath phase, the data time slot starts immediately after the initial transient
overshoot/undershoot period and ends at the point when flow diminishes to an absolute value equivalent to its starting point,
towards the end of the breath phase.
26
Accuracy
±0.5 cm H
O (0.5 hPa) + 4% of set value
2
1
:
±6 L/min or 10% of reading, whichever is greater, at 0 to 150 L/min positive flow
±6 L/min, at 0 to 60 L/min
±30 mL or 20% of reading, whichever is greater
2
±1.0 BPM
2
±20%
Ambient pressure (hPa)
1013.25
956.53
902.41
850.80
801.60
754.73
710.11
O (hPa)
2
O (10 hPa) according to ISO 80601-2-70:2015 and ISO 80601-2-70:2020
2
10 BPM
O [hPa])
0.5
2
O [hPa]): ±1
2
2
2
STPD to BTPS conversion factor
1.12
1.19
1.27
1.36
1.45
1.54
1.65
O (±0.5 hPa)
2
15 BPM
0.5
20 BPM
0.8