Guidance and Manufacturer's Declaration: Electromagnetic Immunity
1188HD Camera is intended for use in the electromagnetic environment specified below.
The customer or the user of 1188HD Camera should ensure that it is used in such an environment.
Immunity
IEC 60601 Test
Test
Level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80MHz to 2.5 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
Compliance
Electromagnetic Environment:
Level
Portable and mobile RF communications
equipment should be used no closer to
any part of the 1188HD Camera system,
including its cables, than the
recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended Separation Distance
3 V
d
=
1.17 P
d
=
1.17 P
3 V/m
d
=
2.33 P
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey
compliance level in each frequency
Interference may occur in the vicinity of
equipment marked with the following
Guidance
80 MHz to 800 MHz
800 MHz to 2.5 GHz
(a)
, should be less than the
(b)
range
.
symbol: