Electromagnetic Compatibility
Like other electrical medical equipment, the 1188HD Camera requires special
precautions to ensure electromagnetic compatibility with other electrical medical
devices. To ensure electromagnetic compatibility (EMC), the 1188HD Camera
must be installed and operated according to the EMC information provided in
this manual.
Note
The 1188HD Camera has been designed and tested to comply
with IEC 60601-1-2 requirements for EMC with other devices.
Caution
Equipment which employs RF communications may affect the
normal function of the 1188HD Camera.
Warning
Do not use cables or accessories other than those provided
with the 1188HD Camera, as this may result in increased
electromagnetic emissions or decreased immunity to such
emissions.
Warning
If the 1188HD Camera is used adjacent to or stacked with
other equipment, observe and verify normal operation of the
1188HD Camera in the configuration in which it will be used
prior to using it in a surgical procedure. Consult the tables
below for guidance in placing the 1188HD Camera.
Guidance and Manufacturer's Declaration: Electromagnetic Emissions
1188HD Camera is intended for use in the electromagnetic environment specified below. The
customer or the user of 1188HD Camera should ensure that it is used in such an environment.
Emissions test
RF emissions CISPR
11
RF emissions CISPR
11
Harmonic emissions
IEC61000-3-2
Voltage Fluctuations/
flicker emissions
IEC61000-3-3
All manuals and user guides at all-guides.com
Compliance
internal function; therefore, its RF emissions are
Group 1
Class A
establishments and those directly connected to
Class A
healthcare professionals only. This system may
Complies
necessary to take mitigation measures, such as
reorienting or relocating the system or shielding
Electromagnetic Environment - guidance
1188HD Camera uses RF energy only for its
very low and are not likely to cause any
interference in nearby electronic equipment.
The 1188HD Camera is suitable for use in all
establishments other than domestic
the public low-voltage power supply network
that supplies buildings used for domestic
purposes, provided the following warning is
heeded:
Warning: This system is intended for use by
cause radio interference or may disrupt the
operation of nearby equipment. It may be
the location.
EN-39