Storage
N/A
Using Sterile Drapes
Using sterile drapes will ensure maximum longevity of your 1188HD Camera
Head. For best results, follow the instructions provided by the drape
manufacturer.
User Maintenance
Replacing the Fuses
1.
Unplug the power cord from the wall outlet and remove the cord from
the camera console.
2.
Unlatch the fuse holder above the AC inlet and remove it. (You may
need to press the tab on the fuse holder with a slender screwdriver to
release the latch.)
3.
Replace the fuse with the same value and rating.
Warning
To avoid the risk of fire, use only fuses of the value specified on
the fuse label located on the rear panel of the camera console.
4.
Reinstall the fuse holder until the tab snaps in place.
Periodic Maintenance Schedule
Warning
To ensure safe operation of the Model 1188HD Medical Video
Camera you should periodically perform the following
procedure:
Every 12 months, check the earth leakage current to <500μA (<300μA in U.S.A.),
ground protective earth impedance to <0.1 ohms, and power consumption less
than or equal to rated power. Use a true RMS digital multimeter and safety
analyzer to perform this test.
Note
Refer calibration and operating difficulties not detailed in this
manual to your Stryker Endoscopy sales representative.
Disposing of the 1188HD
The 1188HD must be disposed of according to local laws and hospital practices.