Guidance and Manufacturer's Declaration: Electromagnetic Immunity
1188HD Camera is intended for use in the electromagnetic environment specified below.
The customer or the user of 1188HD Camera should ensure that it is used in such an environment.
Immunity Test
Electrostatic Discharge
(ESD)
IEC61000-4-2
Electrical fast transient/
burst
IEC61000-4-4
Surge
IEC61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC61000-4-11
Power frequency (50/60Hz)
magnetic field
IEC 61000-4-8
NOTE: Ut is the AC mains voltage prior to application of the test level.
EN-40
All manuals and user guides at all-guides.com
IEC 60601 Test
Level
±6kV contact
±8kV air
±2kV for power
supply lines
±1kV for input/
output lines
±1kV differential
mode
±2kV common
mode
<5% Ut (>95% dip
in Ut) for 0.5 cycle
40% Ut (60% dip in
Ut) for 5 cycles
70% Ut (30% dip in
Ut) for 25 cycles
<5% Ut (>95% dip
in Ut) for 5 sec.
3 A/m
Electromagnetic
Compliance Level
Environment:
Floors should be
wood, concrete, or
ceramic tile. If floors
±2,4,6kV contact
are covered with
synthetic material,
±2,4,8kV air
the relative humidity
should be at least
Mains power quality
should be that of a
±2kV line to ground
typical commercial
±1kV line to line
environment.
Mains power quality
±0.5, 1kV
should be that of a
differential mode
typical commercial
±0.5, 1, 2kV
common mode
environment.
Mains power quality
should be that of a
typical commercial
<5% Ut (>95% dip
in Ut) for 0.5 cycle
environment. If the
user of 1188HD
40% Ut (60% dip in
Camera requires
Ut) for 5 cycles
continued operation
during power mains
70% Ut (30% dip in
interruptions, it is
Ut) for 25 cycles
recommended that
1188HD Camera be
<5% Ut (>95% dip
powered from an
in Ut) for 5 sec.
uninterruptible
power supply or a
Power-frequency
magnetic fields
should be at levels
characteristic of a
N/A
typical location in a
typical commercial
environment.
Guidance
30%.
or hospital
or hospital
or hospital
battery.