Cleaning and Sterilization
The camera console is not intended to come into contact with the patient. It may
be cleaned, but not sterilized. The camera head and coupler may contact the
patient and must both be cleaned and sterilized prior to every use.
Cleaning the Camera Console
Warning
Disconnect the console from the AC power source before
cleaning.
Should the camera console need cleaning, wipe it down with a sterile cloth and
mild cleaning solution.
Caution
Never immerse or sterilize the camera console as this will
damage the camera and void the warranty.
Reprocessing the Camera Head
These reprocessing instructions are provided in accordance with ISO 17664.
While they have been validated by the manufacturer of the medical device as
being capable of preparing the device for re-use, it remains the responsibility of
the processor to ensure that the reprocessing as actually performed, using
equipment, materials, and personnel in the reprocessing facility, achieves the
desired result. This normally requires validation and routine monitoring of the
process.
Warnings
• This device must be cleaned and sterilized prior to the first use and after
every subsequent use.
• Do not sterilize the autoclavable camera head (1188-410-105) with Steris/
Amsco V-Pro or Steris System 1E. Steam sterilization is recommended.
• Use only the sterilization cycles outlined in this document. Using
unspecified sterilization cycles may damage the device or result in
incomplete sterilization.
• Separate the camera head, coupler, and scope prior to cleaning, disinfection,
or sterilization. If the coupler and camera head are cleaned, disinfected, or
sterilized as a single unit, disconnecting the coupler during use will
compromise the sterility of the two products. (Refer to the coupler and scope
product manuals for reprocessing instructions.)
• Wear appropriate protective equipment: gloves, eye protection, etc.
EN-28