Télécharger Imprimer la page

Heidolph Hei-TORQUE Expert Serie Notice D'instructions page 69

Publicité

Les langues disponibles

Les langues disponibles

Troubleshooting
Troubleshooting
Troubleshooting
The following table shows possible faults and solutions:
Error (message)
Display remains dark
after switching on
Rotation cannot be
started
Overload
Overheating
Do not change
during
operation!
Favorite
already exists!
No data
available!
Safety ring is
not fixed
All memory
slots are
assigned!
If a fault cannot be rectified with the described suggestions, please contact an authorized sales
representative or our technical service (see section „Contact information Heidolph international"
on page 76).
Possible cause, remedy
No power supply
Check that the power plug is correctly connected
Electronic defect
Contact technical service
Profile is active
Deactivate profile
Circuit board is defective
Contact technical service
Motor or power supply unit is overloaded
Torque is too high
Blade is blocked
Rotation stops automatically
Remove blockage cause
Switch rotation back on again
Motor too hot: Overheat protection has tripped
Automatic shutdown. Allow the motor to cool and switch
the device back on
Non-permissible change during operation
End process if applicable
Designation already assigned in the system
Rename profile or delete existing profile
In case of Activate profile: Profile data were deleted before
activating or have not yet been entered
In case of Save profile: No profile data entered
Enter profile information
Safety ring not closed correctly
Check the chuck, close the safety ring
Favorites memory is full
Delete no longer required profiles from the system
69

Publicité

loading