Commissioning
The quick-action chuck can be overturned by excessive force applied on
opening! In this case, the quick-action chuck can no longer be closed
normally and a latching noise can be heard on turning it in the closing
direction.
Remedial action: Turn the quick-action chuck again in the opening direc-
tion, beyond the latching point.
The quick-action chuck can then be operated normally again.
Power supply
The device has an appliance outlet with locking system for the power supply. A suitable three-
pole power supply cord is included in the scope of delivery.
DANGER
The device may only be supplied with the required line voltage via a
grounded power socket.
Power cords without a locking system can become loose during operation
due to vibrations. In the event of an uncontrolled shut-down of the stirring
tool, there is a risk of production downtime and or property damage, e.g.
due to uncontrolled restart!
For reasons of operational safety, only use the supplied power supply cord
with locking system!
Connect the power supply cord
➜
Before connecting the power supply cord, make sure that the main switch of the device is
in the OFF position (see following section „Switching the device on/off" on page 58).
➜
Connect the cable coupling of the cord to the IEC appliance inlet on the back of the device.
Make sure that the locking system engages correctly.
➜
Connect the power supply cord to a properly secured mains socket-outlet.
57