Device description
Stirrer on/off swipe field
Stirrer on/off
The [
the switched on device:
Swipe a finger from left to right or from right to left over the field to start and re-stop the
rotational movement.
Starting and stopping the rotational movement does not have any effect
on the device's [
movement merely stops the stirrer tool, the device remains switched on.
The direction of the swipe movement on the swipe field does not have
any influence on the rotational direction of the stirrer tool.
Interfaces
The Hei-TORQUE Ultimate type devices provide two serial interfaces at the back of the
housing:
–
1 × USB Micro type AB
–
1 × RS232 SUB D9
Use these interfaces to connect the device to a PC for external activation.
WARNING
Electric Shock
If the voltages at the connection sockets are too high and insulation is
inadequate, voltage can be applied to metallic parts such as the housing in
case of a fault.
• Extra-low voltage inputs and outputs above 25 V AC or 60 V DC
respectively must be disconnected safely according to EN 61140 or
must be protected by doubled or increased insulation in accordance with
EN 60730-1.
• Only use shielded connection cables and connect the shielding to the
connector housing.
Damage due to penetrating liquid
• Make sure that the interface connections at the back of the device are
secured by the covers provided.
] swipe field is used to start and stop the rotational movement at
on/off
] operating state, i.e. stopping the rotational
52