Operation
➜
Swipe a finger from left to right or from right to left over the field to start and re-stop the
rotational movement.
After a process, stop the rotational movement actively with the swipe
field as described above.
Otherwise the device behaves as described in section „Automatic restart"
on page 58 and possibly restarts automatically!
Enhanced Hei-TORQUE Ultimate functions
The functions described in the following sections are only implemented in
Hei-TORQUE Ultimate type devices!
Timer function
The timer function of the Hei-TORQUE Ultimate type devices can be used to preselect a
required operating period.
Set the timer as follows:
➜
Use the push-and-rotary control to select the
changes to orange).
➜
Press the push-and-rotary control to confirm.
➜
Use the push-and-rotary control to set the required duration:
–
Adjustment Range: 1 min – 24 h
➜
Press the push-and-rotary control to confirm.
➜
The timer is started on starting the rotational movement.
If the timer is running, the preset time is shown on the display to the
nearest minute and the expired time is shown to the nearest second. At
the same time, an alarm clock/timer symbol appears at the top left edge
of the screen.
The timer must be reset to [
The rotational movement can be stopped at any time, even while the
timer is running!
Calibrating the torque
To adjust to the individual environmental conditions on site and for precise display in the
process, the Hei-TORQUE Ultimate type devices provide the option of calibrating the device's
torque. The following options are available:
➜
Relative zero point: the zero point of the torque with immersed stirrer tool is determined
here, i.e., the value results from the resistance of the sample.
➜
Absolute zero point: the zero point on the load-free rotating stirrer tool is determined here.
Before the calibration the device must warm up for 15 – 30 minutes to
compensate for the temperature-dependent resistance of the gear unit.
Timer
option on the Start screen (font color
0
] to deactivate it.
62