Stryker IsoAir 2941 Manuel D'utilisation page 41

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

C C l l e e a a n n i i n n g g
W W A A R R N N I I N N G G
• Always check the CPR connector before use, if visibly soiled, replace. Direct skin contact with soiled material may
increase the risk of infection.
• Always inflate the product and unplug the pump from the mains before you begin cleaning or disinfecting.
• Do not immerse the product in cleaning or disinfectant solutions.
• Do not spray or allow liquids to enter the power cord ports, the hose ports, or the AST socket.
• Always make sure that you wipe each product with clean water and dry each product after cleaning or disinfecting. Some
cleaning and disinfecting agents are corrosive in nature and may cause damage to the product if you use them
improperly. If you do not rinse and dry the product, a corrosive residue may be left on the surface of the product that
could cause premature degradation of critical components. Failure to follow these cleaning or disinfecting instructions
may void your warranty.
C C A A U U T T I I O O N N
• Always completely dry the support surface covers before storing, adding linens, or placing a patient on the surface.
• Do not overexpose the product to higher concentration chemical solutions as these may degrade the product.
• Always follow the manufacturing instructions, failure may affect the useful life of the product.
To clean the support surface and pump:
1. Inflate the support surface with the pump.
2. Unplug the power cord from the hospital grade outlet.
N N o o t t e e - - Leave the hose connected to the support surface and pump to retain the air.
3. Complete steps 4 - 6 for the support surface and repeat for the pump.
4. Wash the outside of the product with a clean, soft cloth, moistened with a mild soap and water solution, to remove
foreign material.
5. Rinse the outside of the product using a clean, soft cloth, and water.
6. Dry the product with a clean, dry cloth.
7. Insert the power cord under the power cord retention flex grip.
8. Plug the pump power cord into a hospital grade outlet.
2941-009-001 Rev AA.0
37
EN

Publicité

Table des Matières
loading

Table des Matières