S S e e t t u u p p t t h h e e s s u u p p p p o o r r t t s s u u r r f f a a c c e e
Unpack the cartons and check all items. Make sure that the product operates before you place into service.
W W A A R R N N I I N N G G
• Always secure the support surface to the bed frame to prevent the shifting of the product while in use.
• Always inspect for foreign objects between the support surface and the bed frame. Foreign objects may cause the
support surface to slide on the support platform.
• Do not use the support surface on a larger or smaller bed frame that does not fit the width, length, or thickness. This is to
avoid the risk of the support surface sliding, patient injury, or interference with moving parts of the bed.
• Do not use the support surface when gaps are present. The risk of entrapment can develop when the support surface is
placed on incompatible bed frames.
C C A A U U T T I I O O N N
• Always be aware of devices or equipment that are placed on the top of the support surface. Damage to the surface may
occur due to the weight of the equipment, heat generated by the equipment, or sharp edges on the equipment.
• Do not put overlays or accessories inside the cover to avoid the risk of reducing pressure redistribution performance.
To setup the support surface:
1. Allow the product to reach room temperature.
2. Lower the siderails to the lowest position of the bed frame.
3. Raise the Fowler to 30 degrees.
4. Remove the headboard and footboard from the bed frame.
5. Loosen the straps that secure the support surface.
6. Unroll and place the support surface with the hose at the foot end of the litter deck. Place the bottom cover face down.
7. Start on one side of the support surface and litter deck. Locate the head end and foot end retaining slot locations on the
litter deck. See the setup for the appropriate bed.
a. Setup with ProCuity bed (page 16)
b. Setup with InTouch bed (page 17)
c. Setup with 3002 S3 bed or 3005 S3 bed (page 18)
d. Setup with GoBed II bed (page 19)
e. Setup with Spirit Plus bed or Spirit Select bed (page 20)
N N o o t t e e
• This product has ten tie down straps. A minimum of four should be used. See the appropriate bed setup for the
number of straps and tie down locations.
• Do not attach the tie down straps to the bed siderails.
8. Thread the tie down strap (E) through the Fowler hole (D) (Figure 2).
EN
12
2941-009-001 Rev AA.0