P P r r o o b b l l e e m m
Disconnected air hose
alarm
Remove from use alarm
Power loss, product does
not turn on
Button not responsive
EN
32
S S c c r r e e e e n n
C C a a u u s s e e
Air hose is disconnected
from the pump or there is a
leak in the system
Software or calibration data
corruption error
Power cord not seated,
power cord unplugged from
outlet
Lock may be on
R R e e c c o o m m m m e e n n d d e e d d a a c c t t i i o o n n
1. Make sure that the air
hoses are seated all the
way into the air hose
ports on the pump.
2. Press the A A c c t t i i o o n n button
next to the A A l l a a r r m m o o f f f f
icon.
3. If the condition persists,
contact a qualified
service technician.
Contact a qualified service
technician.
1. Make sure that the
power cord is plugged
into the product and the
outlet.
2. Secure the power cord
to the flex grip. See
Setup the pump.
3. If the condition persists,
contact a qualified
service technician.
1. Press the A A c c t t i i o o n n button
next to the L L o o c c k k o o n n icon
to unlock the screen.
2. If the condition persists,
contact a qualified
service technician.
2941-009-001 Rev AA.0