S S e e t t u u p p t t h h e e p p u u m m p p
Unpack the cartons and check all items. Make sure that the product operates before you place into service.
W W A A R R N N I I N N G G
• Always plug the power cord into the bed auxiliary outlet or a grounded power respectable. Do not use an extension cord.
• Do not touch patients and the AST socket on the pump at the same time.
• Always secure the pump to the bed frame to prevent the shifting of the product while in use.
• Always check the product before use to make sure there are no broken parts. If the product is damaged call technical
support.
• Always make sure that you do not exceed the safe working load of the bed frame before you mount the pump.
• Always connect the AST cable to the AST socket on the pump to avoid the risk of electrical shock.
• Do not use in the presence of flammable anesthetics, nitrous oxide, or oxygen-rich environments. Risk of explosion can
result.
• Do not expose this product to liquids while powered as this could result in a severe electrical hazard.
• This product radiates radio frequency energy. If not setup and used in accordance with the instructions, may cause
harmful interference to other devices in the vicinity.
C C A A U U T T I I O O N N
• Do not place hoses or power cords in walkways to avoid trip hazards.
• Always allow the product to come to room temperature after exposure to extreme high or low temperatures.
• Always check the power cord for cuts, exposed wires, worn insulation, or other damage that could result in electrical
hazards.
To setup the pump:
1. Allow the product to reach room temperature.
2. Attach the power cord to the pump.
3. Insert the power cord under the power cord retention flex grip (A) (Figure 5).
4. Plug the pump power cord into an accessible hospital grade outlet.
5. Place the pump power cord out of the path of foot traffic to avoid trip hazards.
6. Place the pump on the bed.
a. Setup with ProCuity bed (page 16)
EN
14
A
F F i i g g u u r r e e 5 5 – – C C o o r r d d r r e e t t e e n n t t i i o o n n f f l l e e x x g g r r i i p p
2941-009-001 Rev AA.0