To store the support surface:
1. Start Max Inflate, see Starting or canceling Max Inflate (page 23).
2. Clean and dry the product.
3. Power off the pump.
4. Press the tab in on the CPR connector to disconnect.
5. Remove the CPR connector from the pump.
6. Deflate the support surface.
7. Roll the support surface from the head end to the foot end of the bed.
N N o o t t e e - - Make sure that you insert the AST cable and hose when you roll the support surface.
8. Use the straps on the bottom center to tie the rolled support surface.
N N o o t t e e - - Use the foot end handles to lift and move the support surface for storage.
9. Follow your hospital protocols to store the product.
S S t t o o r r i i n n g g t t h h e e p p u u m m p p
C C A A U U T T I I O O N N - - Always store the product within the specified environmental condition values.
To store the pump:
1. Power off the pump.
2. Unplug the power cord from the hospital grade outlet.
3. Clean and dry the product.
4. Follow your hospital protocols to store the product.